MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MD VIT SLV AND 2.0MM HOMOG CBL; BONE FIXATION CERCLAGE.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MD VIT SLV AND 2.0MM HOMOG CBL; BONE FIXATION CERCLAGE. Back to Search Results
Catalog Number 6704-0-510
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Injury (2348)
Date of Event 05/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding revision involving a dall miles cable was reported.The event was confirmed by medical review.Method & results: product evaluation and results: visual, dimensional, functional inspection, and material analysis were not performed as the item was not returned.Clinician review: a review of the provided medical records and x-rays by a clinical consultant indicated: ¿there is no surgical pathology diagnostic of altr or metallosis, no examination of the explanted components, and no radiographic or clinical evidence of the primary stryker total hip arthroplasty components demonstrating loosening, wear or pathologic changes.After the (b)(6) 2017 revision in which the well-fixed stryker components were replaced with smith & nephew components, all subsequent complications related to the smith & nephew components.At the (b)(6) 2017 revision "no metallosis" was seen and a pre (b)(6) 2017 revision mri was not definitely diagnostic of trunnionosis or pseudotumor and was noted to have implant artifacts.A subsequent (b)(6) 2017 mars mri described a "suspected pseudotumor" at the iliopsoas tendon at a different site from the (b)(6) 2017 description, but also only "suspected".In summary, there is no evidence this clinical picture of complaints of hip pain of six months' duration occurring seven years post-implantation in a patient with chronic low back radicular pain and spine surgery, multiple hip injections and possible chronic hip infection is related to factors associated with the prosthetic hip components.¿ product history review: review of the product history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for this lot.Conclusions: the reported event for revision of the dall miles cable was confirmed.The revision surgery was noted as an unstable right hip for stem subsidence.The stem was a competitor stem which was removed and a larger competitor stem was implanted.The dall miles cables was removed based on the competitor stem¿s removal.
 
Event or Problem Description
It was reported by the patient that she underwent revision surgery of her right hip in (b)(6) 2017.Subsequently she dislocated three times and a spika cast was applied.Stem came out from brace and another revision had to be done.Updated per medical review: ¿on (b)(6) 2017 a revision right total hip arthroplasty (femur only) was performed for an "unstable right tha due to stem subsidence".The operative report notes general anesthesia and use of the previous incision, and during this surgery the s&n revision stem was changed to a larger stem, and uncomplicated surgery was described.¿ updated medical records noted devices were of competitor system, stem and head with 3 stryker cables for the event of subsidence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MD VIT SLV AND 2.0MM HOMOG CBL
Common Device Name
BONE FIXATION CERCLAGE.
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7655382
Report Number0002249697-2018-02011
Device Sequence Number10794569
Product Code JDQ
UDI-Device Identifier07613327036442
UDI-Public07613327036442
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K945294
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number6704-0-510
Device Lot Number57682307
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2018
Initial Report FDA Received Date07/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-