| Catalog Number |
H965457591 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The angiodynamics (b)(6) 2018 complaint report was reviewed for the mini sticks product family and the failure mode "guidewire broke/migrated." no adverse trend was indicated.The hospital's reported complaint description cannot be confirmed; no sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.Possible root cause may be that the guidewire was pulled back against the needle which is cautioned against in the directions for use (dfu).The dfu contains the following statements: "the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire." additionally there is the precaution, " do not advance or withdraw a guidewire against resistance until the cause of the resistance has been determined.Excessive force against resistance may result in guidewire damage or vessel perforation." (b)(4).
|
| |
|
Event or Problem Description
|
|
As reported, regarding ministick, "during attempts at femoral access, the physician attempted to remove the wire and needle to re-attempt.While pulling back, resistance was met and as he began pulling slightly more, the wire tip broke off inside the subcutaneous tissue." no patient complications were reported.The used device was discarded at the hospital.
|
| |
|
Search Alerts/Recalls
|