| Model Number |
DTBA1QQ |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical product: 5076-52 lead, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) rate histograms showed a very high rate of atrial pacing.It was found that the incorrect histogram report was due to a firmware issue whereby a long atrial pace to atrial refractory period can be accounted for as a very short atrial pacing interval.The crt-d remains in use.No patient complications have been reported as a result of this event.
|
| |
|