• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG INSERT 44MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 623-00-44H
Medical Device Problem Codes Mechanical Problem (1384); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 04/09/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding dislocation (closed reduction) involving a trident liner was reported.The event was confirmed by medical review.Method and results: product evaluation and results: visual, dimensional, functional inspection, and material analysis were not performed as the item was not returned.Clinician review: a review of the provided medical records and x-rays by a clinical consultant indicated: ¿when seen on (b)(6) 2016 it was noted she had "dislocated on (b)(6) 2016.Emergency room closed reduction.X-ray slightly displaced fracture anterior wall".On (b)(6) 2016 the office notes states, "no further dislocations.¿ ¿in this obese woman, the primary total hip arthroplasty through an anterolateral approach used a 60 mm shell after reaming to 58 millimeters, which likely compromised the acetabular bone stock.It should be noted that no symptoms referable to the hip are documented prior to the patient's "fell on right side" and developing "deep pain with weightbearing" subsequent to which x-ray revealed "periacetabular fracture of the pelvis." at revision five years later in which acetabular deficiency was noted, the acetabulum was reamed to 65 millimeters for a 66 mm revision component, further compromising acetabular bone stock.Four months later, when the acetabular bone was severely deficient, plate fixation, cage reconstruction and bone grafting was required on a second revision.There is no evidence that factors of component design, manufacturing or materials were responsible for this clinical situation.¿ product history review: review of the product history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for this lot.Conclusions: the event for dislocation (closed reduction) was confirmed.Based on the medical documentation and x-rays provided the patient experienced a dislocation and closed reduction 1 month past revision.The root cause of the dislocation could not be determined.It was noted that there was a fracture on the anterior wall of the acetabulum.As noted there were no further dislocations.Device remains implanted.
 
Event or Problem Description
Per medical review: when seen on (b)(6) 2016 it was noted she had "dislocated on (b)(6) 2016.Emergency room closed reduction.X-ray slightly displaced fracture anterior wall".On (b)(6) 2016 the office note states, "no further dislocations.Pain down posterior leg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG INSERT 44MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7655708
Report Number0002249697-2018-02013
Device Sequence Number13492546
Product Code LPH
UDI-Device Identifier07613327025408
UDI-Public07613327025408
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2017
Device Catalogue Number623-00-44H
Device Lot NumberMLNDAR
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2018
Initial Report FDA Received Date07/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age55 YR
Patient Weight104
-
-