• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X400MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X400MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 34251400S
Health Effect - Clinical Code Fall (1848)
Date of Event 06/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
As reported: "the patient had a long gamma implanted approx.8 years ago at (b)(6).The patient fell last week and the femoral head collapsed.The surgeon decided to remove the gamma nail and do a primary tha using accolade ii size 5 and a trident ha hemi cup 52mm.Routine procedure with no complications".
 
Event or Problem Description
As reported: "the patient had a long gamma implanted approx.8 years ago at (b)(6).The patient fell last week and the femoral head collapsed.The surgeon decided to remove the gamma nail and do a primary tha using accolade ii size 5 and a trident ha hemi cup 52mm.Routine procedure with no complications".
 
Additional Manufacturer Narrative
The reported event that long nail kit r1.5, ti, right gamma3® ø11x400mm x 125° was alleged of 'implant revision / removal without implant breakage' could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The event details did not report an implant failure.The root cause of the event is patient related since the collapse of the femoral head was due to the reported fall.The decision to remove the implant was taken by the surgeon, according to his medical expertise about the case.The implant did not break, and has been functional for the past 8 years.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X400MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key7655716
Report Number0009610622-2018-00254
Device Sequence Number1652522
Product Code HSB
UDI-Device Identifier07613252273851
UDI-Public07613252273851
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2020
Device Catalogue Number34251400S
Device Lot NumberK0A26CE
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/08/2018
Supplement Date Received by Manufacturer07/12/2018
Initial Report FDA Received Date07/02/2018
Supplement Report FDA Received Date08/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-