Model Number 6208-S1
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during an implant procedure both ends of the lead introducer ruptured during cutting.The introducer was not used.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the introducer was returned and analyzed.The analysis indicated that the mechanical operation of the catheter shaft was kinked/buckled.Visual analysis of the lead indicated damage during use.The analyst noted that visual inspection found there was splitting problem on the introducer sheath.The begin of peeling the sheath appeared not on the score line due to a kink, and that caused another side of the handle to break and separated from the sheath.Another side of the handle was not received.The sheath body appeared peeling and cutting.According to the results, the analyst concluded that the sheath kinked causing the handle to break during peeling process.Then the physician tried their best by spiting and peeling the sheath to remove it from the lead.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEELABLE PERCUTANEOUS LEAD INTRODUCERS (PLI)
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key7655791
Report Number2182208-2018-01217
Device Sequence Number10803968
Product Code DYB
UDI-Device Identifier00681490143172
UDI-Public00681490143172
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K911236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2019
Device Model Number6208-S1
Device Catalogue Number6208-S1
Device Lot NumberW4018237
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2018
Initial Date Received by Manufacturer 06/20/2018
Supplement Date Received by Manufacturer09/24/2018
Initial Report FDA Received Date07/02/2018
Supplement Report FDA Received Date10/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
-
-