MAKO SURGICAL CORP. IMPACTOR-ONLAY INSERT; KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED P Back to Search Results
Catalog Number 160405
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It is reported that the impactor has grooves.This led to damaging the onlay.
 
Event or Problem Description
It is reported that the impactor has grooves.This led to damaging the onlay.
 
Additional Manufacturer Narrative
An event regarding damage involving a mako impactor was reported.The event was confirmed by product inspection method & results: product evaluation and results: examination of the returned device with engineer indicated damage observed on the impactor tip consistent with contact against hard object.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Clinician review: a review of the provided medical records by a clinical consultant concluded: product history review: a device history review confirmed all devices accepted into finished goods conformed to specification.Complaint history review: no other events were reported for the lot indicated.Conclusions: the investigation confirmed the reported event as the examination by a material analysis engineer indicated the observed damage is consistent with contact against a hard object.The device was found during surgery but was not used and a different impactor was immediately available.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPACTOR-ONLAY INSERT
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED P
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key7655878
Report Number3005985723-2018-00393
Device Sequence Number16174481
Product Code NPJ
UDI-Device Identifier00848486003203
UDI-Public00848486003203
Combination Product (Y/N)N
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number160405
Device Lot Number12140913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Initial Date Received by Manufacturer 06/06/2018
Supplement Date Received by Manufacturer07/30/2018
Initial Report FDA Received Date07/02/2018
Supplement Report FDA Received Date08/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-