| Catalog Number |
107142 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided pictures showed that the effluent line post pump and the pbp line pre pump were inverted.The reported condition was verified.The cause of the condition was determined to be a manufacturing issue.The involved operator was retrained and awareness training was provided to all operators.The batch involved in this report was manufactured prior to these actions.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the pbp line and the effluent line were found to be crossed on a prismaflex hf1400 set.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|