• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 EM ENT SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 EM ENT SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735669
Device Problems Computer Software Problem (1112); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2018
Event Type  malfunction  
Manufacturer Narrative
Patient age not available from the site.The archive from the procedure was sent to medtronic for evaluation.The archive was found to have points registered below the skin.
 
Event Description
Medtronic received information that, while in a functional endoscopic sinus surgery (fess), the surgeon was unable to track below the eyebrow of the patient.It was noted that the site experienced issues while registering.Adjustment of the patient did not restore functionality.The surgeon opted to complete the procedure without the use of the navigation system.There was a reported delay to the procedure of thirty minutes hour due to this issue.Following the procedure, it was noted that when the emitter was re-positioned, there was lower metal interference from the bed.There was no reported impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
Patient date of birth.A medtronic representative went to the site to test the equipment.Testing revealed that all systems were found to function as expected.The system then passed the system checkout and was found to be fully functional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 EM ENT SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
erika mitchellette
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7656600
MDR Text Key112998239
Report Number1723170-2018-03090
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735669
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age32 YR
Patient Weight112
-
-