COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 840
Medical Device Problem Code Device Inoperable (1663)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, while in use on a patient, the 840 ventilator generated an alarm and generated error codes indicating post (power on self test) failure and oxygen psol (proportional solenoid) valve stuck open and rendering the ventilator to become inoperative.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
 
Additional Manufacturer Narrative
Received additional information on 2018-jul-05 that the correct ventilator serial number was (b)(4).This event was previously reported under mfr report # 8020893-2018-00281.Please reference mfr report # 8020893-2018-00281 for all information regarding this event.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7656699
Report Number8020893-2018-00288
Device Sequence Number1512081
Product Code CBK
UDI-Device Identifier10884521076877
UDI-Public10884521076877
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K151252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840120DICU-EN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/06/2018
Supplement Date Received by Manufacturer07/05/2018
Initial Report FDA Received Date07/02/2018
Supplement Report FDA Received Date07/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-