• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/02/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump was not working.The customer¿s blood glucose level was 212 mg/dl.The customer stated that the insulin pump was exposed to fluid.Troubleshooting was performed but the non-functionality was not resolved.The insulin pump was returned for analysis.
 
Manufacturer Narrative
Unit received with blank display due to corroded electronic assembly.Corroded motor assembly noted during visual inspection.
 
Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0955-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7656773
MDR Text Key113172523
Report Number3004209178-2018-86917
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169656840
UDI-Public(01)00643169656840
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 05/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG1FC7J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/02/2018
Initial Date FDA Received07/02/2018
Supplement Dates Manufacturer Received08/16/2018
10/14/2019
Supplement Dates FDA Received09/12/2018
05/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-0955-2020
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight138 KG
-
-