• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RMT281416
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/19/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.Per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to endoleak and aneurysm enlargement.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2013, the patient presented with an abdominal aortic aneurysm (diameter 55mm) that was treated with the implantation of gore® excluder® aaa endoprostheses.Follow-up imaging performed on (b)(6) 2015, showed aneurysm growth (diameter 58mm).On (b)(6) 2015, a proximal type i endoleak was identified and successfully treated with the implantation of an aortic extender endoprosthesis on (b)(6) 2015.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7657685
Report Number2017233-2018-00363
Device Sequence Number16557242
Product Code MIH
UDI-Device Identifier00733132610723
UDI-Public00733132610723
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2015
Device Catalogue NumberRMT281416
Device Lot Number10870320
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient Weight90
-
-