MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problems
Crack (1135); Device Slipped (1584); Connection Problem (2900)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 06/23/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported by the customer's parent that the customer received emergency medical assistance due to high blood glucose, with unknown blood glucose levels at the time of the incident.The customer's hcp noticed the reservoir ring was loose and there was a crack around the reservoir compartment.The customer was wearing the insulin pump during the incident.It is unknown if the auto mode on the insulin pump was active and was automatically adjusting insulin delivery during the event.Troubleshooting was not completed.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
The test reservoir does not lock in place and unable to perform the displacement test, displacement accuracy test and occlusion test due to missing retainer, broken reservoir tube lip and missing reservoir tube o-ring.Pump passed the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test and force sensor test.
|
|
Manufacturer Narrative
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Z-0958-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|