MEDOS INTERNATIONAL SàRL INTRFX ADV -MOD HEX DRIVER; SURGICAL SCREWDRIVER, REUSABLE Back to Search Results
Catalog Number 254836
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4)¿ incomplete.The expiration date is not currently available.
 
Event or Problem Description
It was reported by the sales rep via phone that during an acl procedure when inserting a 10x30mm screw the tip of their intrafix advance modular hex driver broke off in the head of the screw on the last turn.The tip was unable to be retrieved due to it being lodged tight within the screw.The sales rep stated that the surgeon was fine with it has he stated it would not go anywhere.The procedure had been completed with the device with no harm to the patient or delays to the case.The other part of the device will be returning for evaluation.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.Product complaint # (b)(4).Investigation summary: the complaint device was received and inspected.The complaint was confirmed.It was observed that the distal tip of the hex driver had been sheared off.The distal tip was also twisted.It is possible that the bone hole/screw interaction was abnormally tight causing excessive torque on the device tip therefore causing it to break.However, given the information provided we cannot discern a definitive root cause for the reported failure.A device history record (dhr) review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot dhr review: part number: 254836, supplier lot number: 1706001, qty of lot: (b)(4), release to warehouse date: 9/26/17, manufacturing date: 9/14/17, expiration date: n/a, supplier: (b)(4), manufacturing site: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Udi: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRFX ADV -MOD HEX DRIVER
Common Device Name
SURGICAL SCREWDRIVER, REUSABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 2400
CH  2400
MDR Report Key7658516
Report Number1221934-2018-52849
Device Sequence Number16189036
Product Code LXH
UDI-Device Identifier10886705026296
UDI-Public10886705026296
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number254836
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer07/30/2018
07/30/2018
11/30/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date07/31/2018
07/31/2018
12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-