Model Number BA25-100/I20-40
Medical Device Problem Codes Fracture (1260); Hole In Material (1293); Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in the event will not be returned for evaluation, they remain implanted in the patient.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Event or Problem Description
Patient was initially implanted with a bifurcated, suprarenal, and limb graft in 2015.Patient presented recently with an endoleak type iiib.The physician plans to reline and procedure is scheduled for (b)(6) 2018.No additional patient sequalae has been reported at this time.
 
Event or Problem Description
Patinet secondary procedure has been pushed back from (b)(6) 2018 to (b)(6) 2018.
 
Additional Manufacturer Narrative
At the completion of the investigation, the reported endoleak type iiib of the main body is confirmed.Device, user, procedure or anatomy relatedness of this complaint could not be determined based on the information shared with endologix to review.Procedure related harms for this complaint could not be determined based on the information shared with endologix to review.The final patient status was not reported.There were also evidence to evidence to reasonable suggest strut fractures in the left distal main body stent occurred that was not included in the event as reported.The strut fractures were discovered during review of the 39 month post implant ct scan.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type iiib endoleak.Endologix implemented the following corrective actions with the intent of reducing type iiib endoleak events; upgraded graft material (i.E.Duraply) and updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place (b)(6) 2014.The type iiib endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type iiib endoleak events reported for afx devices has been reduced to less than 0.2%.Root cause investigation was carried out for all afx complaints having an identified failure mode of a type 3b endoleak.Endologix implemented the following corrective actions with the intent of reducing type 3b endoleak events; upgraded graft material (i.E.Duraply) and updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place (b)(6) 2014.The type 3b endoleak rate for afx manufactured and implanted before these corrective actions were put in place is trending at 2.5%.Since the corrective actions were implemented, the type 3b endoleak events reported for afx devices has been reduced to less than 0.2%.The review of manufacturing lot confirmed all devices met specifications prior to release.Devices remain implanted in the patient and were not returned, no evaluation completed.The review of manufacturing lot confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key7658735
Report Number2031527-2018-00579
Device Sequence Number2514729
Product Code MIH
UDI-Device Identifier00818009014108
UDI-Public(01)00818009014108(17)180114
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date01/14/2018
Device Model NumberBA25-100/I20-40
Device Lot Number1310988002
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/03/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-LIMB-LOT: 1253182003; AFX-SUPRARENAL-LOT: 1283063004; AFX-LIMB-LOT: 1253182003; AFX-SUPRARENAL-LOT: 1283063004
Outcome Attributed to Adverse Event Other;
Patient Age68 YR
-
-