| Catalog Number |
PXX41 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a surgical procedure on (b)(6) 2018 and suture was used.During the procedure, the needle was broken after the first passing of the needle-suture through the tissue.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).Date sent to the fda: (b)(4) 2018.Investigation summary a failed suture needle was submitted for evaluation.A fracture was not observed on the needle.The end of the suture attachment was examined in multiple locations and no fracture was observed.
|
| |
|