| Catalog Number |
160245 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
During reduction results surgeon broke prong off distal end of femoral array stem.Prong was located in variable angle screw.Tha procedure.
|
| |
|
Event or Problem Description
|
|
During reduction results surgeon broke prong off distal end of femoral array stem.Prong was located in variable angle screw.Tha procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: it was reported that during reduction results surgeon broke prong off distal end of femoral array stem.Prong was located in variable angle screw.Tha procedure.Product evaluation and results: visual inspection: visual inspection shows broken pin on 160245 array post.Dimensional inspection: dimensional inspection was not completed since visual inspection confirmed broken pin on 160245 array post.Functional inspection: functional inspection was not completed since visual inspection confirmed broken pin on 160245 array post.Product history review: review of the product history records indicate 51 devices were manufactured under lot no w29586.And accepted into final stock on 11/12/ 2012.No non-conformances were identified during inspection.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 160245, lot w29586.Shows no additional complaints related to the failure in this investigation.Conclusions: visual inspection confirms broken pin on160245 array post.Functional inspection was not completed since visual inspection confirmed broken pin on 160245 array post.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
|
| |
|