• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ATTUNE REV P-F STEM TRL 18X60; STEM TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ATTUNE REV P-F STEM TRL 18X60; STEM TRIALS Back to Search Results
Catalog Number 251318060
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 06/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during revision the stem drill broke off the broach.The stem trial was stuck in the patients tibia and it was retrieved without having any cause or damage to patient.All pieces were retrieved from the patient.Surgical delay of 20-30 minutes.
 
Additional Manufacturer Narrative
(b)(4).Examination of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV P-F STEM TRL 18X60
Common Device Name
STEM TRIALS
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key7658879
Report Number1818910-2018-63712
Device Sequence Number1108646
Product Code LXH
UDI-Device Identifier10603295438892
UDI-Public10603295438892
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number251318060
Device Lot NumberNW214346
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Initial Date Received by Manufacturer 06/08/2018
Supplement Date Received by Manufacturer09/24/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
-
-