| Catalog Number |
109.1006 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that nearly 4 months after the dental implant and abutment were placed in ada site 16 the mouth, the implant had not yet achieved osseointegration in type ii bone.The product will be forwarded to the manufacturer for investigation.There were no reported complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
|
| |
|
Event or Problem Description
|
|
The clinician reported that nearly 4 months after the dental implant and abutment were placed in ada site 16 the mouth, the implant had not yet achieved osseointegration in type ii bone.The product will be forwarded to the manufacturer for investigation.There were no reported complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number: e2015015.(b)(4).Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
|
| |
|
Event or Problem Description
|
|
The clinician reported that nearly 4 months after the dental implant and abutment were placed in ada site 16 the mouth, the implant had not yet achieved osseointegration in type ii bone.The product will be forwarded to the manufacturer for investigation.There were no reported complications.(b)(4).
|
| |
|