The suspected lot number was either h18a29114 or h17l06038.Five actual samples were received for evaluation without pouches; therefore unable to differentiate the units to the reported lot numbers.A batch review was conducted for each potential lot and there were no deviations found related to this reported condition during the manufacture of either lot.A visual inspection with naked eye noted a hole/cut in the heater bag on one of the amia cassettes.Functional testing, including leak testing, clear passage testing, and clamp function testing was performed; leak testing failed due to the hole/cut in the heater bag.The reported issue was verified.The cause of the condition could not be determined.The other four samples passed all testing and met specifications.Should additional relevant information become available, a supplemental report will be submitted.
|