Model Number 8830415003
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, while performing a catheter insertion procedure, the kit's guide wire could not pass through the tip of the catheter's lumen.It stated that there was a blockage inside the lumen.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted no abnormalities.Pmv performed functional testing, the guide wire was inserted and removed and reinserted into the tube without restriction.The catheter was submerged into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.A review of the dhr (device history record) indicates the product was released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during insertion, the kit's guide wire could not pass through the tip of the catheter's lumen.It was stated that there seemed to be a blockage inside both venous and arterial lumen.It was also stated that the doctor has accessed the internal jugular vein for insertion and local anesthesia was given.The vein was dilated with kit's dilator and then the guidewire was inserted thru the catheter lumen which was not passing through reflecting some blockage in the passage of lumen.The procedure was completed with another catheter kit.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key7658947
Report Number3009211636-2018-00221
Device Sequence Number2063499
Product Code FOZ
UDI-Device Identifier20884521005171
UDI-Public20884521005171
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K943349
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2021
Device Model Number8830415003
Device Catalogue Number8830415003
Device Lot Number1619700144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Initial Date Received by Manufacturer 06/14/2018
Supplement Date Received by Manufacturer08/22/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date09/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-