| Model Number |
8886621741 |
| Medical Device Problem Codes |
Break (1069); Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic gynecologic procedure, the needle broke and fell into the patient's cavity and was retrieved.A new device was used to complete the case.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the opened samples noted one suture and one needle.The needle was received broken.Upon microscopic inspection, instrument marks were observed near the break site.Additionally, both broken ends of the needle parts show twist marks.As no sealed samples were returned, a bend moment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.A definitive root cause could not be determined with regard to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|