| Catalog Number |
6215-5-011 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
When completed, the evaluation summary will be submitted in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Two separate bone plugs were opened for surgery and deemed unsterile due to seal on packaging.Update (b)(6) 2018: spoke to rep.The seal on the outer blister was not adhered to the outer blister all the way around and there were 'gaps' allowing air to pass in and out.The seal on the inner blister was also not adhered to the inner blister all the way around with the same 'gaps'.This happened with 2 devices with 2 different lots.The outer boxes were discarded before the rep could retrieve them.Primary tka procedure.Rep was present in the or and witnessed the event.Additional devices were immediately available in the or and there was no surgical delay - the surgeon was not even aware of the event until after the procedure was complete.
|
| |
|
Event or Problem Description
|
|
Two separate bone plugs were opened for surgery and deemed unsterile due to seal on packaging.Update 12/jun/2018: spoke to rep.The seal on the outer blister was not adhered to the outer blister all the way around and there were 'gaps' allowing air to pass in and out.The seal on the inner blister was also not adhered to the inner blister all the way around with the same 'gaps'.This happened with 2 devices with 2 different lots.The outer boxes were discarded before the rep could retrieve them.Primary tka procedure.Rep was present in the or and witnessed the event.Additional devices were immediately available in the or and there was no surgical delay - the surgeon was not even aware of the event until after the procedure was complete.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding pack damage involving a bone plug was reported.The event was confirmed.Visual inspection: an inspection of the returned device noted the outer seal was opened.The packaging was otherwise unremarkable.Medical records received and evaluation: not performed as there was no patient involvement.Review indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Review indicated there have been no other events for the reported lot.It was deemed through nc that this event is not manufacturing related.The conclusion of this nc stated there are sufficient controls and inspection steps in the manufacturing process to verify seal integrity and that the shrinkwrap meets all workmanship standards.It can be concluded that the seal breaches of the complaint parts did not occur during the manufacturing process, and must have occurred externally.The exact cause of the event could not be determined because insufficient information was provided.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|