BAUSCH + LOMB STELLARIS VISION ENHANCEMENT SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number BL2380
Device Problems Aspiration Issue (2883); Unexpected Shutdown (4019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2018
Event Type  Injury  
Manufacturer Narrative
The compressor, functioned as required during testing.A complete functional test was performed and compressor passed all checks and met all of the manufactures specifications.(b)(4) did not exhibit or show evidence of the complaint issue.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLARIS VISION ENHANCEMENT SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer Contact
juli moore
3365 tree court industrial blv
st. louis, MO 63122
6362263220
MDR Report Key7659213
MDR Text Key119965049
Report Number0001920664-2018-00077
Device Sequence Number1
Product Code HQC
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K063331
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberBL2380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/03/2018
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-