• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 16F; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 16F; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 000716
Medical Device Problem Codes Burst Container or Vessel (1074); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 06/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter allegedly fell out of the patient.Reportedly, there was a narrow gap between the stomach wall and the abdominal wall.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Investigation summary: the sample returned for evaluation was one 16 fr tri-funnel replacement gastrostomy feeding tube.Visual, functional, and microscopic evaluations were performed of the physical sample.The sample was returned with a longitudinally-aligned c-shaped split in the balloon material.The surrounding material near the split site had a distended shape memory and dark usage discoloration.The observed damage is consistent with a burst due to over-pressurization.Therefore, the investigation is unconfirmed for 2923e - feeding tube dislodged and confirmed for 1074a - balloon ruptured.The investigation indicates that the failure is resulting from balloon rupture, and while a definitive root cause could not be determined based upon available information, procedural issues may have contributed to this event.Labeling review: a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, possible complications and contraindications showed that the product labeling is adequate.(expiry date: 06/2021).
 
Event or Problem Description
It was reported that the catheter allegedly fell out of the patient.Reportedly, there was a narrow gap between the stomach wall and the abdominal wall.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 16F
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7659220
Report Number3006260740-2018-01595
Device Sequence Number16187030
Product Code KNT
UDI-Device Identifier00801741037139
UDI-Public(01)00801741037139
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000716
Device Catalogue Number000716
Device Lot NumberNGCN0404
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Initial Date Received by Manufacturer 06/11/2018
Supplement Date Received by Manufacturer03/12/2019
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date03/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-