• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC TURBO-JECT POWER INJECTABLE; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TURBO-JECT POWER INJECTABLE; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER Back to Search Results
Model Number G56194
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Pma/510(k) #: k111244.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported the wire would not pull our of the catheter after a turbo-ject power injectable picc line was inserted.The patient was said to have angulated and tortuous vessels.Another similar device was used to complete the procedure successfully.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Event or Problem Description
See h10.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information has previously been reported, is unknown or is unavailable.The complaint device was returned on (b)(6) 2018 and evaluated on (b)(6) 2018.The investigation was completed on (b)(6) 2018 and reviewed on (b)(6) 2018.The returned used device met specification and did not show signs of fraying, unraveling, or fragmentation/separation.This complaint no longer meets the fda 21 cfr part 803 definition of a device malfunction.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT POWER INJECTABLE
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7659287
Report Number1820334-2018-01900
Device Sequence Number945236
Product Code LJS
UDI-Device Identifier00827002561946
UDI-Public(01)00827002561946(17)200223(10)8616147
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
K111244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2020
Device Model NumberG56194
Device Catalogue NumberUPICDS-5.0-CT-OTW-ST-1110
Device Lot Number8616147
Was Device Available for Evaluation? Yes
Event Location Hospital
Initial Date Received by Manufacturer 06/15/2018
Supplement Date Received by Manufacturer10/05/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date10/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-