The complaint receiver device was returned for evaluation.The device was externally visually inspected and no defects were found.The receiver failed to charge and reboot.The receiver did not turn on.The battery was replaced with a known good battery and the receiver turned on.The receiver log was downloaded, rttloss was observed.The receiver case was opened, the internal inspection failed with a damaged battery controller chip.Functional testing was unable to be performed because the receiver had been opened.The reported event of unexpected receiver shutdown was confirmed.The root cause was determined to be a damaged battery.
|