• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C1650
Medical Device Problem Codes Bent (1059); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The stretch resistance wire (sr wire) was fractured, the embolization coil was detached from its pusher assembly inside its introducer sheath, and the proximal constraint sphere was inside the distal detachment tip (ddt).The introducer sheath was ovalized approximately 6.0 cm from the distal tip.Conclusions: evaluation of the returned ruby coil revealed that the introducer sheath was ovalized.If the device is retracted out of its packaging hoop at extreme angles, damage such as ovalization may occur.Further evaluation of the returned device revealed that the embolization coil sr wire was fractured.If the embolization coil is forcefully retracted against resistance, damage such as this may occur.The embolization coil being retracted through the damage introducer sheath likely contributed to the sr wire fracture.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a coil embolization procedure, the hospital staff found that the introducer sheath of the ruby coil was bent upon removal from the packaging.The damage to the introducer sheath was found prior to use.Therefore, the ruby coil was not used in the procedure.The procedure was completed using a new ruby coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7659498
Report Number3005168196-2018-01313
Device Sequence Number1146391
Product Code HCG
UDI-Device Identifier00814548013244
UDI-Public00814548013244
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2025
Device Catalogue NumberRBY4C1650
Device Lot NumberF77224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2018
Initial Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-