• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information not provided due to (b)(6) patient privacy regulations.Unique device identification (udi) unavailable.Device manufacturing date is unavailable.A medtronic representative went to the site to test the equipment.The representative reported that the computer was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.
 
Event Description
Medtronic received information regarding a navigation device being used for an unknown procedure.It was reported that while using the device, it switched to "no input detected".The site rebooted the device a few times without resolution.They switched the power outlets but no resolution.Eventually, the device was able to boot up for the case.There was no delay to surgery and no impact on patient outcome.
 
Manufacturer Narrative
Unique device identification (udi) and device manufacture date provided.Product and related fields updated to proper value.The computer for the navigation system was returned to the manufacturer for evaluation.Testing found that the hard drive was removed upon receipt.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7659626
MDR Text Key113258971
Report Number1723170-2018-03085
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9735665
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/07/2018
Initial Date FDA Received07/03/2018
Supplement Dates Manufacturer Received07/09/2018
Supplement Dates FDA Received07/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-