• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. EVOLUTION STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. EVOLUTION STERILIZER Back to Search Results
Device Problem Fluid Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2018
Event Type  Malfunction  
Manufacturer Narrative

A steris service technician arrived onsite following the reported event. While inspecting the unit, the technician found a crack in the water hose and a worn funnel, resulting in the reported water leak. To resolve the issue, the technician replaced the water hose and funnel, ran a test cycle, and found the unit to be operational. The sterilizer was returned to service.

 
Event Description

The user facility reported water leaking from their evolution sterilizer. No injury or procedure delay or cancellation was reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEVOLUTION STERILIZER
Type of DeviceEVOLUTION STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7659889
MDR Text Key113304175
Report Number3005899764-2018-00054
Device Sequence Number1
Product Code FLE
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial
Report Date 07/03/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/03/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/05/2018
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-