| Catalog Number |
M0036121450 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The subject device is not available.
|
| |
|
Event or Problem Description
|
|
During the coil embolization procedure, it was reported that the coil was prematurely detached inside the patient when it was being reheated.Therefore, the physician used a retriever device to remove the detached coil from the patient.The procedure was aborted and then resumed the other day.No clinical consequences reported to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The subject device was not returned; therefore, physical as well as a functional evaluation could not be performed.Information available indicated that the device was prepared as per the dfu (direction for use) and was in a good condition prior to use.Based on the information currently available the exact cause for the reported event cannot be determined.
|
| |
|
Event or Problem Description
|
|
During the coil embolization procedure, it was reported that the coil was prematurely detached inside the patient when it was being resheathed.Therefore, the physician used a retriever device to remove the detached coil from the patient.The procedure was aborted and then resumed the other day.No clinical consequences reported to the patient.
|
| |
|