• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSTEC BARBADOS INC. SOFTEC I; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSTEC BARBADOS INC. SOFTEC I; INTRAOCULAR LENS Back to Search Results
Lot Number 174364
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lens was returned in the vial, the cap was on the vial but the bung was absent.The outer pouch was also absent but the remainder of the packaging components were present.Inspection was performed at 10x magnification under a stereoscopic microscope.The lens was intact and one haptic carried a scratch.No other defects were noted.The dimension check conducted showed that the lens was within specification.A full audit of all batch documentation relating to the production of the devices was performed.The audit concluded that all procedures in manufacturing and packaging of the devices had been conducted correctly.Batch reconciliation was 100.0%.Based on the assessment of our batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the devices were conducted correctly.The lens was intact and within specification.There was no evidence to suggest that any aspect of the lens was responsible for the surgical complication reported.[(b)(4) pdf].
 
Event or Problem Description
Lenstec received and email stating "dr.Had trouble centering the lens so she took it out".The implant/explant date was (b)(6) 2018.The incision was enlarged to remove the lens and there was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC I
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church, BB170 92
BB  BB17092
MDR Report Key7660507
Report Number9613160-2018-00029
Device Sequence Number13739703
Product Code HQL
UDI-Device Identifier00844369029605
UDI-Public00844369029605
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date01/10/2023
Device Lot Number174364
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/07/2018
Initial Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-