(b)(4).Manufacturing site evaluation: there is a clear visible charring at the right side of the upper jaw.A visual inspection of the jaw was performed.Except the charring no further abnormities was found.Then the mechanical functions we checked.The clamping and the cutting function worked well.The manufacturing documents have been checked and found to be according to specification which was valid during the time of the production.One probable root cause in cases like this is a not a proper cleaning during surgery, so the carbonized residues of blood or tissue initiate are arc.A further possible root cause is a damaged insulation tongue (dissection clip).This damage created by wrong handling during cleaning the electrodes.A damage during the rf-generator failures can be excluded.There is suspicion of a design related aspect to the failure, this failure mode has been into a capa.
|