| Model Number |
H74939293051530 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of the event: 18 years or older.(b)(4).Device evaluated by mfr: device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed. bsc id: (b)(4).Tw#: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that stent foreshortening occurred.A 5x150x130 innova¿ stent was selected for an intervention in the patient's superficial femoral artery (sfa).The lesion was 90% stenosed, minimally tortuous, and moderately calcified.Antegade access was gained to the lesion and pre-dilatation was performed.There was no difficulty deploying the stent, however, during deployment the stent started to compress and it shortened.The stent measured 80 mm post-deployment (visual estimation).No damage was noted on the stent.Post-dilatation was performed after the stent was deployed.Because the stent was assessed to be shortened, a 5x60 innova¿ stent was used to cover the rest of the lesion.No patient complications were reported.The patients status after the procedure was fine.
|
| |
|