Model Number 8225101
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No evaluation completed as the device has not been returned at this time.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A distributor reported that the probe was functioning properly at the beginning of the surgery.The probe was unable to detect nerves after about 30 minutes into the surgery.There was no warning sound when the probe was near nerves.The doctor reconnected the probe but the problem was not resolved.The doctor then switched to a new probe and the issue resolved.The thyroid surgery was delayed for about 5 minutes.It was noted that there were no patient symptoms or complications associated with the event.
 
Additional Manufacturer Narrative
Device evaluation determined that visually, there was no damage or anomalies in the construction of the device that would have resulted in the reported event.The tip fit securely in the handle.The resistance of the assembly shall measure less than 0.3 ohms while using a 4-wire ohm meter (a 2-wire meter was used in lieu of the 4 wire); the actual measurement was 0.2 ohm.There was no intermittent behavior while manipulating the tip, connector, and wire.The probe was plugged into a nim system (with a patient simulator) using 1.0 ma stimulating current and 100uv threshold; when stimulating, the system returned 1.02 ma and a waveform / event tone over 300uv.There was no out of specification condition identified as it relates to the complaint.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® PROBE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7660926
Report Number1045254-2018-00277
Device Sequence Number2629284
Product Code ETN
UDI-Device Identifier20681490176686
UDI-Public20681490176686
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K934426
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/26/2025
Device Model Number8225101
Device Catalogue Number8225101
Device Lot Number0214481040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer07/11/2018
08/08/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date08/02/2018
10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-