| Model Number |
15 |
| Medical Device Problem Code |
Device Inoperable (1663)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer¿s device and verified the reported issue.It was also observed that one side of the power button was difficult to activate.Physio replaced the device's main keypad assembly.The battery pins were also replaced as a preventative measure.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that intermittently their device would have a blank display and the service indicator would be illuminated.In this state defibrillation therapy may not be available to a patient if needed.There was no patient use reported for this event.Upon evaluation of the customer¿s device, physio-control observed that the device was unable to complete a boot cycle and had a blank display.
|
| |
|