• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227); No Code Available (3191)
Date of Event 05/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
Reportedly, the patient has complained since her first post-op day that she has a blurred haze around all objects which is intensified by light.The date of implant was (b)(6) 2018.The patient¿s vision was 20/15 at distance and j1 at near.The doctor reported that the symfony intraocular lens (iol) is centered perfectly, but appears to have circular swirl marks on the anterior surface of the central optic, which seem to be perfectly symmetrical and could not have been done by human hands.Per follow-up, the lens was explanted (b)(6) 2018 and replaced with the same model, a smaller diopter size lens.Patient was doing fine post-surgery.No further information was provided.
 
Additional Manufacturer Narrative
The same diopter size, 21.5 of replacement lens was utilized rather than smaller size, as initially reported on mdr (medical device report).Device available for evaluation? yes.Returned to manufacturer on: 8/1/2017.Device returned to manufacturer? yes.The product was received in a lens case.Visual inspection with the unaided eye shows that the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The reported complaint cannot be confirmed.The manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7661563
Report Number9614546-2018-00638
Device Sequence Number1108598
Product Code POE
UDI-Device Identifier05050474605237
UDI-Public(01)05050474605237(17)210920
Combination Product (Y/N)Y
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/20/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/20/2021
Device Model NumberZXT150
Device Catalogue NumberZXT150U215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2017
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/07/2018
Supplement Date Received by Manufacturer08/28/2018
Initial Report FDA Received Date07/03/2018
Supplement Report FDA Received Date09/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Date Report Sent to FDA01/01/2005
-
-