• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC POWERED SURGICAL SOLUTIONS LEGEND TAPERED DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC POWERED SURGICAL SOLUTIONS LEGEND TAPERED DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number F1/8TA15
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that both tools returned has similar fracture locations near the fluted end portion.The likely cause was identified as the fracture location, orientation and surface texture is consistent with tools breaking when a side load was applied on the tool during use.Bending stress caused the tool to bend and break.Tools were most likely not rotating when each broke.The device user manual warnings section includes the following warning ¿do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.¿ we will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that two of the same type of tools were broken when a suture hole of craniotomy line was being created.It was reported that there was no patient impact.No further information was obtained upon follow up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND TAPERED DISSECTING TOOL
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MEDTRONIC POWERED SURGICAL SOLUTIONS
4620 north beach street
ft. worth TX 76137
MDR Report Key7661706
Report Number1625507-2018-00032
Device Sequence Number2629955
Product Code HBE
UDI-Device Identifier00643169766235
UDI-Public00643169766235
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2023
Device Model NumberF1/8TA15
Device Catalogue NumberF1/8TA15
Device Lot Number0214621481
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2018
Initial Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-