| Model Number |
MMT-551NAH |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was over 300 mg/dl at the time of the incident.The customer was assisted with troubleshooting and customer reports display was blank currently.Customer states the display did not returned.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The customer was advised the insulin pump will be replaced as per troubleshoot.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump was received with blank display due to moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Unable to perform operating currents, self-test, rewind test and displacement test due to blank display.
|
| |
|