| Model Number |
Lag Screw |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported to the company that tapping was not performed prior to lag screw insertion into the bone.As described in the system ifu and surgical technique, tapping should be performed before lag screw insertion into the bone, using the provided bone tap.It is noted that the operation was satisfactorily completed.Examination of the production records of the involved device did not reveal a deviation that may have contributed to the reported case.
|
| |
|
Event or Problem Description
|
|
A piccolo composite pf lag screw was damaged during implantation.Another screw was used and the operation was satisfactorily completed.
|
| |
|