Model Number MMT-523NAB
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Unit received unable to perform the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery due to complete broken reservoir tube lip and end cap broken off noted and also received with cracked battery tube threads, cracked lcd window, minor scratched lcd window, missing reservoir tube o-ring, cracked case display window corner, missing end cap sticker, stained address/serial number label, broken belt clip slot and cracked case reservoir tube window corner.The p-cap not locks into the reservoir compartment properly due to completely broken reservoir tube lip noted.
 
Event or Problem Description
Customer reported via phone call that insulin pump had bottom of pump where the drive support cap.Customer's blood glucose value was unknown at the time of the incident.The customer was assisted with troubleshooting.Customer will return device for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7662271
Report Number3004209178-2018-87891
Device Sequence Number1979549
Product Code OYC
UDI-Device Identifier00643169202306
UDI-Public(01)00643169202306
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAB
Device Catalogue NumberMMT-523NAB
Device Lot NumberA8523NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2018
Initial Report FDA Received Date07/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight190
-
-