| Catalog Number |
K09A |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively primary osteoporosis and underwent balloon kyphoplasty.During surgery, the contrast medium leaked when the balloon was inflated in the vertebral body at l2.The product was taken out of the body and being checked, it seemed that there was a crack in the balloon.Ibt ruptured during inflation.The patient was not allergic to the contrast media.The product came in contact with the patient.No patient complications were reported as the result of the event.There was a delay of less than 60 minutes.Patient's issue was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: visual and microscopic evaluation revealed that the ibt was damaged.The balloon had large radial damage to the balloon sheathing, one balloon was cut completely into and was separated from the shaft.This extensive damage is consistent with the balloon being damage from being pulled back through the cannula after a rupture has occurred.Once the material on the balloon has been breached the balloon loses its ability to become pressurized and its ability to hold a negative pressure which aids in the removal process.The balloons were likely cut in the vertebral body then damaged more on the removal process.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|