CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD Back to Search Results
Model Number 7742
Device Problems Capturing Problem (2891); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2018
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this patient's heart rate decreased and his right ventricular (rv) lead exhibited high pacing thresholds and failed the automatic threshold testing a few days after it was implanted.Further analysis revealed the rv lead was pulled back.As result, the patent underwent a revision to resolve the issue.The lead remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7663050
MDR Text Key113154371
Report Number2124215-2018-12054
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/07/2020
Device Model Number7742
Other Device ID NumberINGEVITY MRI
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7741; 7742; L311
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age64 YR
-
-