• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. D/B/A EV3 NEUROVASCULAR AXIUM PGLA 3D; DEVICE, EMBOLIZATION, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. D/B/A EV3 NEUROVASCULAR AXIUM PGLA 3D; DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number PC-2-6-3D
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the coils did not detach.The patient was undergoing a vessel take down surgery to treat a tumor.It was stated that the first coil did not detach.As a result a second coil of the same size and lot number were used which also failed to detach.The healthcare professional (hcp) attempted to detach the coil 4 times with the instant detacher and 2 times with a manual method, with no other instant detacher attempted, but the issue was unresolved.No issue with the instant detacher was noted.There were no related patient symptoms.Additional information received noted that the patient had a normal vessel tortuosity.The patient was ultimately treated with a different coil and are currently in good condition.There was no damage to the pushwire noted once it was removed from the patient, and a continuous flush was administered during the procedure.
 
Additional Manufacturer Narrative
The axium coil was returned for evaluation inside of a (b)(6) shipping box and with the opened axium coil inner pouch.The axium coil was found to be inside of a biohazard ziploc bag.The instant detacher was not returned.During evaluation, the axium coil was removed from the biohazard ziploc bag.The tack weld replacement (twr) crimps and the pushwire distal to the break indicator at the manual detachment location (via hypotube) were found to be intact.There was an appearance of blood on the pushwire and within the proximal end of the introducer sheath.The implant coil appeared to be damaged and stretched within the introducer sheath.The axium coil could not be pushed out from within the introducer sheath for further evaluation as resistance was encountered.The axium coil was pulled out from within the introducer sheath.The axium implant coil was found to be damaged and stretched with the polypropylene filament intact.An in-house instant detacher was then selected for detachment testing.The axium implant coil successfully detached from the pushwire on the first attempt.All other subassemblies appeared to be normal and no other anomalies were observed.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PGLA 3D
Common Device Name
DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. D/B/A EV3 NEUROVASCULAR
9775 toledo way
irvine CA 92618
MDR Report Key7663642
Report Number2029214-2018-00601
Device Sequence Number1524735
Product Code KRD
UDI-Device Identifier00847536014145
UDI-Public00847536014145
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/05/2019
Device Model NumberPC-2-6-3D
Device Catalogue NumberPC-2-6-3D
Device Lot NumberA304911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2018
Supplement Date Received by Manufacturer09/14/2018
Initial Report FDA Received Date07/05/2018
Supplement Report FDA Received Date10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age32 YR
-
-