Model Number H7493926028350
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Codes Chest Pain (1776); Myocardial Infarction (1969); Thrombosis (2100)
Date of Event 05/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is a combination product.Method: labeling evaluation.The device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is anticipated procedural complication as the event is due to a known physiological effect of the procedure noted within the directions for use, and/or device labeling.(b)(4).
 
Event or Problem Description
It was reported via facility medwatch # 5077592 that stent thrombosis occurred.In (b)(6) 2017, the patient presented to emergency room (er) with st-elevation myocardial infarction (stemi) and was then admitted.The patient was sent to cardiac catheterization lab which revealed total occlusion in the mid right coronary artery (rca) with large amount of thrombus.Overlapping inflations was then performed using 2.5x15 emerge balloon catheter providing sluggish flow.Thrombectomy was performed and brisk flow returned with minimal residual defect.Subsequently, a 3.5x28mm synergy stent was deployed in the rca with good results.The procedure was completed with no patient complications.The patient did well and was discharged on acetylsalicylic acid (asa) and brilinta.In (b)(6) 2018, the patient experienced chest pain and called emergency medical services (ems).Patient takes asa and brilinta before arriving in the er with stemi.Cardiac catheterization revealed total thrombotic occlusion of the previously placed stent in the mid rca which was then treated with thrombectomy and 3.5x20 nc emerge balloon.Timi iii flow was achieved and the procedure was completed.The patient did well and was discharged on asa and prasugrel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7663867
Report Number2134265-2018-05883
Device Sequence Number19464668
Product Code NIQ
UDI-Device Identifier08714729840305
UDI-Public08714729840305
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/02/2018
Device Model NumberH7493926028350
Device Catalogue Number39260-2835
Device Lot Number0020618609
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2018
Initial Report FDA Received Date07/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-