Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained of 532, 575, 424, 456 and 389mg/dl.The expected fasting blood glucose test result range is 150 to 200mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a back to back blood test was performed fasting by the customer and produced test results of 389 and 456mg/dl using true track meter.The test strip lot manufacturer's expiration date is 07/19/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly).(b)(6).
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained of 532, 575, 424, 456 and 389mg/dl.The expected fasting blood glucose test result range is 150 to 200mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a back to back blood test was performed fasting by the customer and produced test results of 389 and 456mg/dl using true track meter.The test strip lot manufacturer's expiration date is 07/19/2020 and open vial date is june 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) result 1 :532mg/dl, date: on (b)(6) 2018, time:9:06 am, fasting, result 2 :575mg/dl, date: on (b)(6) 2018, time:9:04 am, fasting, result 3 :424mg/dl, date: on (b)(6) 2018, time:9:01 am, fasting, result 4 :164mg/dl, date: on (b)(6) 2018, time:1:06 am, fasting, result 5 :193mg/dl, date: on (b)(6) 2018, time:9:56 pm, fasting.
|