| Model Number |
MICL13.7 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lens was returned in liquid, in lens vial.Visual inspection found the optic torn and portion of the lens missing.One similar complaint type event was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.7mm micl13.7, -13.0 diopter, implantable collamer lens tore during loading.There was no patient contact and the backup lens was implanted with no issues.Cause of event was unknown.
|
| |
|
Search Alerts/Recalls
|