Catalog Number 222344
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/10/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Evaluation statement: the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed 1 similar complaint for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the doctor used a 3.4 healix peek anchor (ref # (b)(4), lot # 3686346) in tandem with a single implant transtend system for a pasta repair.He drilled and tapped all the way to the laser line for both anchors.The single loaded anchor that came with the transtend set went in without any problems.The first 3.4 healix peek anchor stripped approximately three quarters of the way down.They drilled down on top of the anchor and then attempted to set another anchor on top of it, which broke as it was being screwed in.They were able to unscrew it and remove the anchor, and then put in the third and final anchor, which took and was used in the repair.The sales rep confirmed there were no patient consequences and a ten minute delay.Bone quality was good not hard.The sales rep could not provide further patient information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX PEEK 3.4MM
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7664343
Report Number1221934-2018-51200
Device Sequence Number11202427
Product Code MAI
UDI-Device Identifier10886705023837
UDI-Public10886705023837
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/10/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2016
Device Catalogue Number222344
Device Lot Number3686346
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/10/2013
Initial Report FDA Received Date07/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-