BOSTON SCIENTIFIC - COSTA RICA (COYOL) ULTRASONIC IMAGING CATHETER - CORONARY; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number UNK97
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
 
Event Description
Same case as 2134265-2018-06331 and 2134265-2018-06041.It was reported that automatic pullback failure occurred.An ilab ultrasound imaging system was used in conjunction with an imaging catheter and a pullback sled; however, it froze while it was being used.During auto pullback, upon pulling the motor drive about 5 to 6cm a reboot message was displayed.When ok was clicked, the image suddenly turned blue and it became unable to manipulate.The ilab system was checked onsite and the software was upgraded.After this, pullback functional check was performed more than 10 times with no problem.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC IMAGING CATHETER - CORONARY
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7664428
MDR Text Key113250758
Report Number2134265-2018-06332
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNK97
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-