• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1737
Device Problems Detachment Of Device Component (1104); Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/19/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, part of the insulation tip came off the jaw and fell apart.There was no patient injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
This event has been reassessed.The event is no longer considered a serious injury, and is now classified as a malfunction.Evaluation summary: one device sample was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product did not meet specification as received.Visual inspection found the bottom jaw over mold was damaged and missing plastic.The disengaged plastic was not returned.The investigation found the bottom jaw over mold was damaged and missing plastic near the hinge of the device.The damage to the jaw's over mold likely occurred during the insertion or extraction of the device jaw from a trocar instrument.The investigation identified the root cause of the reported event to be user error.The instruction for use (ifu) states, carefully insert and withdraw the instrument through the cannula to avoid damage to the device and or injury to the patient.Close jaws using device handle before insertion/extraction in the trocar.Corrective actions have been implemented to mitigate this issue.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
lisa hernandez
5920 longbow drive
boulder, CO 80301
2034925563
MDR Report Key7664629
MDR Text Key113254754
Report Number1717344-2018-00900
Device Sequence Number1
Product Code GEI
UDI-Device Identifier20884521181578
UDI-Public20884521181578
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K141153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/07/2022
Device Model NumberLF1737
Device Catalogue NumberLF1737
Device Lot Number73100115V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-